Taking measurements

The Measurement section of the sidebar holds Line, Circle, TPA, CORA, Norberg, Tangent (the common tangent), Q angle (the quadriceps angle), Axis (the anatomic axis), and THR cup and THR stem (the two halves of a total hip replacement). Two lines that cross also read the angle between them.

Line, Circle, TPA, THR cup and THR stem need a calibrated radiograph. CORA, Norberg, Q angle and Axis read only angles, which is why they work on a film with no scale set. The common tangent’s angle works without one too; its advancement is in millimetres and needs a scale. All of them scale automatically with zoom and rotation.

Measurement modes

How do I measure a length?

Distance between two points. Used for fragment length, plate span, bone width, and any general linear measurement.

Arm Line and click the film once: a line drops there, starting where you clicked. Drag each end onto its landmark. Dragging on the film instead of clicking moves the film, not the line.

How do I measure an angle?

Draw two lines that cross, one along each axis or edge, each from landmark to landmark. Where they cross, the film reads the angle between them in degrees. The lines must actually cross on the film, so lengthen one if they do not reach — lock its angle first and it cannot tip while you do.

Two crossing lines make four angles. Click inside the one you want and it becomes the reading, or click the number to move it round to the next. With either line selected, the sidebar offers the angle and the one beside it, acute or obtuse. The choice is saved with the case.

The angle is listed in the Items list under the second of the two lines, and printed in the surgical plan. Each line also reads its own length, so lines need a calibrated radiograph.

How do I lengthen a line without changing its angle?

Select the line and press Lock angle, under its length in the sidebar. Its ends then slide along the line, so dragging either one makes it longer or shorter and nothing else, however your hand wanders — the way to stretch a line until it crosses another without disturbing the angle. An anatomic axis has no lock: its ends always slide along it. The arrow keys move a locked end along the line too.

A padlock beside the length on the film shows the line is locked. It cannot be moved by its middle while it is; press Lock angle again to free it. The lock is saved with the case.

How do I draw an anatomic axis?

The anatomic axis runs through the middle of a bone’s shaft, centred between its two cortices at two levels. Press Axis. On a view with one long bone, such as a side view of a tibia, the axis is placed straight away. On a view with more, such as a ventrodorsal view of the hips, point at the bone you want: the axis snaps to it, and a click places it. Either way it runs from one end of the bone to the other, with a short dashed line across the bone at each level, which the axis runs through the middle of. For another, such as the other femur, press Axis again and click that bone. Off a bone, a click places one for you to drag where you want it.

The levels sit at shares of the axis’s length, measured from the end that starts at the top. Its panel offers four pairs in one press — 33%, 50% (the femur’s anatomic axis, which aLPFA and aLDFA are measured from), 25%, 50%, 33%, 67%, and 50%, 85% (the tibia’s distal anatomical axis, which the anatomical-mechanical axis (AMA) angle is measured from) — and two boxes to type any others. It starts at 33%, 50%, or at 50%, 85% on a view with a TPA.

Its ends are placed on the ends of the bone: for a femur, the top of the femoral head and the stifle. If an end is not on the end of the bone, drag it there. An end slides along the axis, and the levels follow to their shares of the new length and centre themselves on the cortices again, so its Length is the bone’s. Each level stays square to the axis, with a grip at each end. If the end of a level has missed the cortex, drag it onto the cortex: the level follows it up or down the bone, and the axis re-centres through the level’s new middle.

Where the axis crosses a line, the two read the angle between them. To carry it to a line it does not reach, drag its end past the line: the end only slides along the axis.

How do I draw a mechanical axis?

Draw a line from one joint centre to the other: for the femur, from the centre of the femoral head to the middle of the intercondylar notch; for the tibia, from the middle of the intercondylar eminence to the middle of the talus. To find the centre of the femoral head, fit a circle to it first — its radius ends on the centre. Where the mechanical axis crosses another line, the two read the angle between them.

How do I measure femoral varus (aLDFA)?

On a craniocaudal view of the femur, press Axis (and click the femur, if the view shows more than one bone): the axis is placed through the middle of the femoral shaft a third and half way down, from the top of the femoral head to the stifle. Check its ends sit on the ends of the femur and drag them there if not. Draw a line across the distal femoral joint, along the distal condyles; if the axis does not reach it, drag its lower end down to meet it. Where they cross, the film reads the angle; click inside the angle on the lateral side to read that one. That is the aLDFA. The aLPFA is read the same way, against a line from the tip of the greater trochanter through the centre of the femoral head.

On a CT scan, pressing Femur in Measurement reads the aLDFA, the aLPFA as well, and the mechanical angles mLDFA and mLPFA, measured against the line from the centre of the femoral head to the middle of the knee. Click a measurement’s box to draw it, with its handles: the head’s centre, the trochanter’s tip and the knee’s middle can each be dragged. It also shows the femur’s bend: front-on as varus or valgus, and side-on as procurvatum or recurvatum. That is the angle between the axes of the femur’s upper and lower parts, the same figure a CORA dropped on the femur reads. No normal range is shown beside it, because a normal femur is not straight. Pressing Tibia shows the tibia’s bend the same way.

How do I measure a femur, tibia, radius or stifle on a CT scan?

On a CT scan, Measurement has a button for each part of the leg — Femur, Tibia, Radius, Hip, Stifle and Elbow — above the drawing tools. One press measures everything for that part the scan shows:

  • Femur: the aLDFA, the aLPFA, the mechanical angles mLDFA and mLPFA, the neck-shaft angle (NSA), the version, the femur’s bend front-on and side-on, and its length.
  • Tibia: the TPA with the mCaPTA, the mMPTA, the mMDTA and mCrDTA at the hock, the tibial torsion, the tibia’s bend and its length.
  • Radius: the aMPRA and aCdPRA at the elbow, the aLDRA and aCdDRA at the carpus, the radial torsion, the radius’s bend front-on and side-on, and its length. On a scan of both forelimbs, both radii are measured. Check their axes and joint lines on the scan, and drag a handle where one is off.
  • Stifle: the TPA, the TT-TG, the sulcus angle and depth, the Q angle, and the patellar tilt, which you place yourself: press its box, then click.
  • Hip: for each hip, how the acetabulum covers the femoral head: the DASA, the CEA, the dorsal cover and the acetabular version. The next question has what each one is.
  • Elbow: the radioulnar step and the humeroulnar and humeroradial joint spaces.

Each measurement has a box in the panel. One is drawn on the scan at a time: click a box to draw that one instead, with its handles to adjust, and the views turn to where it is measured. A box that stays empty says why when you point at it; press it to try that measurement on its own, then click on the scan to place it yourself if it still can’t (the hip’s are read off the bone, so they cannot be placed by a click). Where a press can place nothing at all, it is ready for that click straight away. The stifle and the tibia share one TPA. On a scan of both legs, Left and Right above the boxes choose the leg, and each length says how it compares with the other leg’s; the first time a stifle or tibia measurement is pressed with no leg chosen, the app asks which, with Left and Right on the message, and that leg stays chosen for the scan. One undo takes back everything a press placed. In Items, each press is one row, its measurements folded under it.

How do I measure hip dysplasia on a CT scan?

On a CT scan of the hips, press Hip in Measurement. For each hip it reads how the acetabulum covers the femoral head, on the slice through the head’s centre square to the pelvis’s long axis:

  • DASA, the dorsal acetabular sector angle: from the line between the two femoral heads round to the dorsal rim of the acetabulum. A lower figure means less cover.
  • CEA, the centre-edge angle: from the line straight up from the head’s centre to the same rim, positive where the rim reaches past it. Read to the bony rim, it is the DASA less 90°.
  • Dorsal cover: the share of the upper half of the femoral head that the acetabulum covers, counted all round the head. It is the same figure the DPO/TPO in Osteotomy reads before its turn.
  • Version: how far the opening of the acetabulum faces down as well as out, from the line between its dorsal and ventral rims. A lower figure means an opening facing more upwards.

Each box shows both hips, right first. Click a box to draw that measurement on the scan. The figures are measured against the pelvis’s own planes, not the scanner’s. A CT shows the hip lying still, not how loose it is, so there is no Norberg angle on a CT. On a scan with one femur, the DASA and CEA are read from the pelvis’s midline instead of the line between the heads.

How do I measure the anatomical-mechanical axis (AMA) angle?

The angle between the tibia’s anatomic and mechanical axes, on a side view: how far the top of the tibia tips backwards. It reads higher in dogs prone to cruciate rupture.

On a mediolateral view of the tibia, press Axis. With a TPA already on the view its levels start half way down and just above the distal flare (50%, 85%); otherwise press 50%, 85% in its panel. Check its ends sit on the ends of the tibia. Draw a line along the tibia’s mechanical axis, from the middle of the intercondylar eminence to the middle of the talus. Where the two cross, the film reads the AMA angle — drag the axis’s end to lengthen it if they do not meet.

How do I measure a circle?

Diameter, radius or circumference of a circular feature — joint surfaces, medullary canals, or bone fragments.

Arm Circle and click once: a circle drops, centred where you clicked, with a handle at the top and one at the bottom. Drag each handle onto opposite sides of what you are measuring: the circle runs between them, and the handle you are not dragging stays put. Drag inside the circle to move it. It reads the radius, the diameter and the circumference — in full in the sidebar, and as R, Ø and C on the film.

A radius runs from the circle’s top handle to its centre, and ends exactly on the centre. Move that handle to turn the radius. Press Lock radius in the sidebar and its direction holds: either handle then only makes the circle bigger or smaller along that line, without turning it. Where a line crosses the radius, the two read the angle between them, just as two crossing lines do.

TPA (tibial plateau angle)

A two-line construction used in TPLO planning: the mechanical axis of the tibia, and the tibial plateau line. The TPA reading appears in degrees, just past the caudal end of the plateau line, so it stays off the joint whichever way the film faces.

On a mediolateral view that reaches the talus, pressing TPA places it on the four points for you, ready to adjust. On one that stops short of the talus, its axis is estimated down the middle of the shaft, and its panel says so — see TPLO planning.

Refer to your standard TPLO planning references for target angles.

Where is the CORA for a CBLO?

It is an option on the TPA. Select the TPA and turn on CORA in its panel, and two lines join it. A solid line runs up the middle of the tibia, from the marrow cavity halfway down, through Gerdy’s tubercle — where a TTA cut starts — and 15 mm past it. It is placed for you, with a handle at each end to drag. Where the tubercle cannot be found, the line runs through the middle of the marrow cavity a third of the way down instead. A dashed line, the proximal axis, runs from the top of the TPA’s axis at 90° less the target TPA to the plateau: 80° for the usual CBLO target of 10°. Where the two lines cross is the centre of rotation a CBLO is planned from. A short dashed line labelled TBL, the transverse bisecting line, crosses the bone through that point.

The panel reads the Magnitude, the angle measured between the two lines. It is not the CBLO’s turn: a CBLO turns by whatever lands the plateau on the target TPA, which can differ from the magnitude. The Target TPA starts at 10°, the usual target for a CBLO, whatever your TPLO default in Settings; type another and press Enter. It is the CBLO’s target too — change it here or in the CBLO’s panel, and both show the same figure, the CORA moves, and a CBLO planned from it moves with it. If the app cannot place the shaft line, it starts on the TPA’s own axis and the panel asks you to drag its two handles onto the marrow cavity and Gerdy’s tubercle. Turn the option off to take both lines away. The Surgical plan and the plan summary on the film list the magnitude and its target TPA under the TPA.

CORA

Centre of rotation of angulation. Draw the proximal and distal axes; the construct reads the deformity angle and marks the CORA where the axes cross.

Pressing CORA places one on the main long bone of the film for you, ready to adjust: a joint centre on each joint, each axis drawn from its joint centre through the middle of its half of the shaft, and the CORA where the two axes cross. If the bone has no bend to find, the CORA is placed halfway along it, reading 1°, and a note says so: move its handles to the bend you see. If it finds no bone on the film, a note says so and you place the CORA by clicking, as before. If the film already has a CORA, pressing CORA again lets you place another by clicking, as before: a second bend is placed with a second CORA. Pressing CORA only ever places one on the view you have selected, so to put a CORA on the view beside it, select that view first.

The angle is the whole bend between the two axes as drawn. Where a normal bone is not straight in that view — a radius on a lateral, with its cranial bow — type that bow into the Normal bow box on the wedge’s panel, not the CORA’s: the wedge then takes out the angle less the normal bow; see Why does the wedge aimed from a CORA look like too much?

Can the CORA’s axes run down the middle of the shaft?

Yes. Turn on Anatomic axes in the CORA’s panel. Each axis then runs through the middle of its own part of the shaft, centred between the cortices at two levels: two levels in each half of the shaft, found on the bone for you, and moved to either side of the bend where it sits nearer one end. With the switch off, each axis runs from its joint centre through the shaft, as before. A new CORA starts with it off.

Each level stays square to its axis, with a grip at each end on the cortex. Drag a level’s end onto the cortex: the level follows it up or down the bone, and its axis re-centres through the level’s new middle. Each axis runs on through the CORA until it leaves the bone. Drag an axis’s end to slide it along the axis. If the bone cannot be read, the levels sit on the axes as they were, and the panel asks you to drag their ends onto the cortices. Turn the switch off to put the axes back where they were.

The two kinds of axes usually read different angles on the same bone, so the Surgical plan says when a CORA used anatomic axes. A normal bow typed on a wedge has to be measured the same way as the CORA it is aimed from.

Norberg

The Norberg angle, both hips at once. Drop both femoral head centres and both craniolateral acetabular rims; it reads each hip live, with 105° or more shown as normal.

On a ventrodorsal view of the hips with the hips extended, pressing Norberg places it for you — both head circles and both rims — ready to adjust. If the hips are small on screen, the film zooms in on them. If another view is selected, it goes on the hip film, and that film becomes the selected one. On any other film, or if you click first, it drops where you click as before.

Each head centre is found from a circle: drag its three handles onto the edge of the femoral head, and the centre follows. Until you touch one, the second head’s circle copies the size of the first. If you would rather lay one ring over a head and size it, as with a transparency, press Ring for that hip in the panel.

The L and R labels follow the patient, not the screen. A lead side marker on the film, where the app has read one, decides which hip is which. Without one, the app goes by how the view is usually hung, and mirroring the radiograph tells it the film had been stored back to front. Flipping the film top to bottom never changes the labels — see How do I flip an X-ray?.

THR

A total hip replacement is planned in two parts, each a tool of its own: THR cup, the acetabular component, set against the pelvis, and THR stem, the femoral component, set against the femur. Both need a calibrated film, because they read in millimetres. The implants themselves — cups and stems — go on the same film from the Implants tab, at true size, like any other template. The neck cut is its own tool, FHNO on Osteotomy — see The FHNO.

On a ventrodorsal view of the hips, press THR cup, point at the hip you want and click. The app reads both hips and fits the cup to the one nearer your click: the acetabulum’s cranial edge, caudal edge and medial wall, the inside of the medial wall, the dorsal rim, and the pelvic midline, laid through the two femoral heads. If the hip is small on screen, the film zooms in on it. It is selected and ready to adjust, every mark a handle you can drag, and one undo takes it back. The midline is drawn only while the cup is selected.

Press THR stem and click the same hip for the femur’s half: the greater trochanter, the trochanteric fossa, and the canal’s width at the lesser trochanter and at the isthmus. A canal width the film did not show is drawn dashed: drag its two handles onto the inside of the cortices.

How do I plan only the cup, or only the stem?

Place just the one you need: THR cup and THR stem are separate tools, and each lists, reads and prints on its own. Both can sit on the same hip. A THR planned before they were split still shows both halves as one.

How do I plan the stem on a lateral view, aiming for the patella?

Press THR stem on a lateral view of the femur and click the femur. The app finds the bone and draws the proximal femoral axis through the middle of the shaft a quarter and half way down, with the entry where it crosses the intertrochanteric fossa, the femoral condyles and the front of the patella — drag any of them to adjust. Patella is the patella’s angle off that axis, seen from the entry. Aim is where the axis passes the patella: aim the reamers and broaches that far cranial or caudal of the patella to follow the canal. Take the view with the stifle at about 90°, as it is held in theatre: there the patella sat about 1.6° caudal of the axis in the one study that measured it, so aiming straight at it nearly follows the canal. It moves forward as the stifle straightens and back as it bends.

How do I check a THR after surgery?

Tick Post-op at the top right of the post-op film — it shows on every film while THR cup or THR stem is pressed — then press either and click the implant:

  • THR cup places a Cup check. On a lateral view, put five points on the cup’s ventral margin and the line’s ends on the ilial wings and the ischiatic tuberosities: it reads the cup’s opening and the version of its open face, as Fink and Lotsikas (2026) read them. On a ventrodorsal view of a cemented cup, put the five points on its wire ring: it reads the opening alone. That view cannot give the version, and under 45° it reads the opening high (Dyce and others 2001), so read those on a lateral view; a cup with no wire ring is read on a lateral view too. Switch between the two under Read on.
  • THR stem places a Stem check. On a ventrodorsal or caudocranial view it reads the stem’s shoulder below the greater trochanter’s tip, which is what subsidence is measured from on later films, and the stem’s angle, varus or valgus. On a lateral view it reads where the neck points: the stem’s version from the neck against the stem, for the neck angle you type (135° until you do). The femoral condyles and the patella tell it which way is cranial.

A figure outside the published target shows amber; hover a tile for the target and its source. There is no validated post-op score for canine hips, so each check reads one figure the papers measure. The Surgical plan prints them all.

What is the cup on the THR?

A faint dotted circle through the three acetabular points is the acetabulum as measured, and Acetabulum is its diameter. The planned cup is a solid-edged shape drawn over it: a three-dimensional cup, projected onto the film as the beam would see it, so its shell, its open face and its cut show as what they are. Under Planned cup in the panel:

  • Truncated or Hemispherical: a cup with its dorsal rim cut away at 45° to its open face and a 3 mm cylindrical rim, or a plain half-ball. The cut face is hatched, and the heavy line is its edge.
  • Cup size: one of the sizes the makers offer, starting at the one nearest the acetabulum. The cup keeps the acetabulum’s centre, so a bigger cup beds deeper.
  • Opening: the angle of lateral opening, 0° facing the floor and 90° straight out. It starts at 45°; the system you are planning with says what it should be.
  • Version: the cup’s open face against the pelvic midline, positive retroverted. Until you set it, the face lies along the acetabulum’s dorsal rim, the dashed line between the cranial and caudal points. With the opening, it sets which way the cup faces.
  • Inclination: turns the cup on its own axis, as you would on the inserter, carrying the cut round the rim while the cup keeps facing the same way. At 0° the cut’s edge is level; positive turns its caudal end dorsally. A hemispherical cup has no cut, so no inclination.
  • Cut edge: under the sliders, the cut edge’s own version and inclination, as the papers measure a truncated cup.

Each slider changes the cup on the film as you drag it, with one undo per drag, and each has a box to type in. Drag the cup itself, or its centre handle, to slide it towards the medial wall; its centre is the hip centre the plan reconstructs.

What do the THR’s figures mean?

On the THR cup: Reamed in is how far the cup’s bed is set past the floor of the socket, from sliding the cup in and from a cup bigger than the acetabulum; below zero, in amber, the cup is smaller than the acetabulum and its bed stops short of the floor. Wall left is the bone left between the bed and the inside of the medial wall on this view, amber below zero. Rim cover is how much of the cup’s width the dorsal rim covers. Rim angle is the dorsal rim against the midline, a guide for the version you set.

On the THR stem: Isthmus is the canal’s width at its narrowest, read square across the femur — the width a stem has to pass. Flare index is the canal’s width at the lesser trochanter over its width at the isthmus: under 1.8 is a stovepipe femur and shows amber, over 2.5 a champagne flute. Both read a dash while a width is still a guess. Trochanter is where the greater trochanter’s medial edge sits against the canal’s lateral wall, amber when it is medial of it and may stand in a straight stem’s way. Entry is where the reamer goes in, at the trochanteric fossa, against the femoral axis; the dashed path on the film runs from there to the isthmus.

Hover a tile for the full definition and the source it follows. The Surgical plan prints the cup’s size, opening, version and inclination, and every figure.

How do I measure the common tangent for a TTA?

Press Tangent on a mediolateral view of the stifle, ideally with the stifle at about 135°. On a side view of the knee it is placed for you, with no click on the film: a circle on the femoral condyles, a circle on the tibial condyles, the cranial edge of the patella at its distal end, and the patellar ligament’s insertion on the tibial tuberosity. On any other film, click the stifle joint to place it.

Either way, drag its handles to fit the bone: three on each circle’s outline, the patella and the insertion. If the two condyles do not line up, put each circle halfway between the two outlines. The tangent runs at right angles to the line between the two circles’ centres, where the circles meet.

It reads the PTA, the angle between the patellar ligament and the common tangent, and the Advancement: how far the tibial tuberosity has to come forward, along the tangent, for the PTA to be 90°.

How do I measure the Q angle for patellar luxation?

The Q angle (quadriceps angle) estimates the pull that luxates the patella: the line of the quadriceps against the line of the patellar ligament. It is measured on bone landmarks alone; the muscles themselves cannot be seen well enough on a film or a CT to say more.

Use a ventrodorsal view that runs from the hips to below the stifles, with the limbs pulled straight and the patellae centred. Press Q angle, then click on the stifle. Drag its three points onto:

  • the rectus femoris origin, at the cranial margin of the acetabulum;
  • the middle of the trochlea, where the patella sits in its groove;
  • the tibial tuberosity, where the patellar ligament inserts.

The dashed line carries the quadriceps line on past the trochlea, and the angle is how far the ligament turns away from it. It needs no scale. In healthy dogs on ventrodorsal films the median is about 18° under 15 kg and 9° over 15 kg (Pinna and Romagnoli 2017). It is higher with medial patellar luxation, but it does not grade a luxation on its own, so nothing turns amber.

On a CT scan of the hip and stifle, pressing Stifle in Measurement places it for you, with the stifle’s other measurements. The origin is read off the hip and the groove and the tuberosity off the stifle. The angle is read as a ventrodorsal film would show it: in the femur’s frontal plane, with the stifle straightened to 160°, about as straight as a dog’s stifle goes, so it does not depend on how the dog lay in the scanner. A scan without the hip cannot give it. The femur’s shaft does not show where the hip is, so it is not estimated: the TT-TG (under Stifle) measures where the tuberosity sits, and the aLDFA (under Femur) how the end of the femur is angled.

Is a measurement accurate if the calibration is only estimated?

An angle is. Norberg, CORA, the TPA, the angle between two lines and the common tangent’s PTA are worked out from where the points sit on the film, not from its scale, so they read the same whether the scale is estimated, measured off a ball, or — for Norberg, CORA and the anatomic axis — not set at all. What decides an angle’s accuracy is where the points are.

A length is only as good as the scale. With Estimated showing beside the px/mm figure, the scale is the DICOM’s own pixel size, and a Line, a Circle’s diameter, an advancement in millimetres and an implant’s size can be 10–25% out. Calibrate off a ball before trusting those — see What does “Estimated” beside the scale mean?.

How do I place the points accurately?

  • Magnification loupe. Toggle it from the top bar for precise endpoint placement on bony landmarks.
  • Snapping. When you place a new measurement, clicking within a few pixels of an existing endpoint or centre starts it exactly on that point, and a ring shows when it will. Dragging an endpoint afterwards does not snap.
  • Arrow keys. Rest the mouse on a point — or drag it roughly into place and let go — then tap the arrow keys to walk that point one pixel at a time (Shift + arrow for ten). Its ring turns heavier while the arrow keys are holding it.

How do I hide the measurements on the film?

The ruler button in the top bar hides every reading on the film at once — the millimetres and degrees — if they are cluttering the view; the lines stay. On a narrow window it is Readings on the film in the menu. To take a measurement’s lines off the picture as well, use the eye on its row in the Items list. Nothing is deleted either way, and the surgical plan still lists them.

Do measurements stay on the film?

  • They remain on the radiograph until you delete them, and the case auto-saves.
  • They live in image coordinates, not screen coordinates — they zoom, pan, and rotate with the radiograph.
  • As many as you like can sit on the same radiograph.

How do I move or delete a measurement?

  • Move an endpoint: click and drag the small handle at either end.
  • Move the whole measurement: click and drag the line body. A line whose angle is locked moves only by its ends.
  • Move its number out of the way: drag the reading itself. It stays where you put it through zoom, rotation and flipping, and when the case is opened again, with a dotted line back to what it measures. Double-click it to put it back.
  • Delete: select and press Delete or Backspace. On a phone or tablet, press and hold it until a red bin appears beside it, then tap the bin.